Designed to maximize turnaround time, RAPIDLab® 1265 Systems have the capacity to enhance blood gas testing efficiency in the clinical laboratory by combining testing speed and reliability with low maintenance. RAPIDLab 1265 systems are our proven solution for high-volume critical care testing sites.
Designed to meet the challenges of the busy clinical lab or various hospital settings, RAPIDLab® 1265 Systems address all your critical care testing needs:
- Extensive test menu includes full CO-oximetry and neonatal total bilirubin
- Ready Sensor® electrode technology for industry-proven reliability
- Cost-effective, cartridge-based system
- Simplified operation with biosafe sampling
- Automatic quality control and documentation
- Seamless integration with your LIS/HIS
- Dedicated technical support and on-site service
Enhanced critical care testing in the NICU
RAPIDLab 1265 Systems’ testing performance, accuracy and consistency make it an excellent choice as your NICU monitoring solution:
- Ultra-fast sample processing enables immediate physician intervention
- Clinical accuracy reduces the need to repeat tests
- Microsample capability enables testing of very small sample volumes without compromising accuracy
- Full test menu from a single sample
When it comes to such small and vulnerable patients, fast results, delivered instantly, can make a big difference.
Get a complete picture in just 1 minute
- RAPIDLab 1265 Systems parameter listing:
Blood Gas (pH, pCO2, pO2)
Electrolytes (Na+, K+, Ca++, Cl– )
Metabolites (Glucose, Lactate, Neonatal Total Bilirubin)
CO-oximetry (tHb, HHb, O2Hb, sO2, COHb, MetHb)
- Testing panels can be customized to meet specific needs
- Capable of running the full test menu on short-sample patient draws (microsamples)
Increase your throughput
- Intuitive touchscreen and automatic single sample port (syringe and capillary) speed up the testing process
- All reagents are contained in sealed cartridges for enhanced operator safety and low maintenance
- Clot detection and clearance with biosafe automatic sampling system
Get lab-quality results and help ensure regulatory compliance
- Onboard automatic quality control cartridge requires no operator intervention during the 28-day use life
- The entire sample pathway is analyzed at an operator-definable frequency
- Long-life Ready Sensor® electrodes ensure industry-proven accuracy and reliability
Increase control and connectivity
- Seamless integration with our RAPIDComm® Data Management System or third-party vendor connectivity solutions
- Laboratory managers/supervisors have direct access to monitor and control all Siemens critical care systems on the hospital network
- Instant clinician access to patient results from networked POC analyzers
- Dual LIS transmission support enables both the serial port and the ethernet port to communicate simultaneously
Parameters* | Analyte |
---|---|
Blood Gas | pH pCO2 pO2 |
Electrolytes | Na+ K+ Ca++ CI– |
Metabolites | Glucose Lactate |
CO-oximetry | tHb Neonatal Total Bilirubin |
*Assays vary by model
Sample Size: Syringe (1 mL)/Capillary (Minimum) | RAPIDLab 1240 | RAPIDLab 1245 | RAPIDLab 1260 | RAPIDLab 1265 | |
---|---|---|---|---|---|
pH Only | 35 µl | 35 µl | 95 µl | 95 µl | |
CO-oximetry Only | — | 100 µl | — | 100 µl | |
Full Menu | 90 µl | 140 µl | 125 µl | 175 µl | |
Analysis | |||||
Time to Results: | 60 seconds | ||||
Cartridges | |||||
Reagent Cartridge Use Life: | 30 days | ||||
Wash Cartridge Use Life: | Sample dependent | ||||
AQC Cartridge Use Life: | 28 days | ||||
Calibration | |||||
1 point Calibration | Every 30 minutes | ||||
Full Cal Calibration | Every 8 hours | ||||
Environmental | |||||
Temperature: | 15°C-32°C | ||||
Humidity: | 5%-85% non-condensing | ||||
System Dimensions | |||||
Width: | 58.4 cm (23.0 in.) | ||||
Height: | 57.2 cm (22.5 in.) | ||||
Depth: | 55.9 cm (22 in.) | ||||
Weight: | 29.5 kg (65 lb) | ||||
Power Requirement | |||||
Rating: | 150VA | ||||
Voltage: | 100-240V | ||||
Frequency: | 50/60Hz |
External Interfaces |
---|
USB Printer Port, RS232 Port, Bar Code Scanner Port, RJ10/100 Base T Ethernet Port |
Approvals |
---|
EN 60601-1-2:2001, Group 1, Class B, CB Test Certificate, UL Certificate, CSA Certified |